AE-1152D_11C DC Low Ohm METER
Optimum for the automatic measurement of super low-resistance resistor.Select the result of judgment by maximum 11 classifications with BIN function.
- Optimum for super low-resistance resistor etc. of shunt resistor.
- Cancelling the influence of thermal electromotive force.
- Impressed the pulse measurement current to reduce self-heating and the abrasion of measurement terminal.
- Contact-check is equipped as standard function.
- The measurement of the absolute value :0.0000mΩ~1.5000kΩ
- %measurement :0.1mΩ~1kΩ/±50.00%【The minimum resolution 10nΩ】
- GP-IB/RS-232C/Centronics interface.【optional】
- BIN function[BIN1~BIN9]【Available for maximum 11 classifications】Available for open-collector output,LED display and buzzer.
- Equipped with check circuit as standard function to find out whether the measurement current/measurement voltage is abnormal.
- Available for shift output.
Measuring range and Accuracy (at 23℃±5℃):
Range | Measuring range | Resolution | Measuring current | Accuracy |
---|---|---|---|---|
1mΩ | 0.0000mΩ~1.5000mΩ | 0.1μΩ | 3A | within ±(0.01%rdg +1μΩ) ±3digit 【Average】 ±4digit 【Slow】 ±5digit 【Fast】 |
10mΩ | 0.000mΩ~15.000mΩ | 1μΩ | 1A | |
100mΩ | 0.00mΩ~150.00mΩ | 10μΩ | ||
1Ω | 0.0000Ω~1.5000Ω | 100μΩ | 100mA | |
10Ω | 0.000Ω~15.000Ω | 1mΩ | ||
100Ω | 0.00Ω~150.00Ω | 10mΩ | 10mA | |
1KΩ | 0.0000kΩ~1.5000kΩ | 100mΩ | 1mA | |
% | 0.1mΩ~1kΩ/±50.00% | 0.01%[10nΩ] | Refer below | Refer to the instruction manual |
Voltage at terminals open : | below 5V. |
---|---|
Measuring method : | 4-terminal measurement[Available for contact -check.] |
Measuring time : | 【Free running mode】2~10times/sec. |
【External start mode】about 9msec~400msec. | |
Comparator setting range : | [0~15000][ for both the upper and lower limit、%range:low 0~-50%、high 0~+50% |
Indication comparator’s comparison result : | [BIN1~BIN9] display and buzzer. |
Control signal : | Start:”L”[0V]→”H”[DC12V] |
Hold:Open and H”[DC12V]:Free run、”L”[0V]:Hold | |
Signal of judgment result [LO/GO(BIN1~BIN9)/H1] :open collect output max40V、100mA |
|
Contact error[CONT-E]: open collect output max40V、100mA | |
completion signal [EOC]:open collect output max40V、100mA[40V/100mA max.] | |
Operation condition : | Temperature:5℃~+40℃、 Humanity:Below 85% |
Outer dimension : | 333(W)×99(H)×300(D)mm【excluding protruding parts such as rubber legs, etc.】 |
Weight : | about 3kg |
Power supply : | AC85V~265V、50~60Hz、about 60VA |
Option | ・RS-232C | Only one kind of interface can be provided with the checker. |
---|---|---|
・GP-IB | ||
・Printer output[Centronics] | ||
・Printer Cable | ||
・Short termination(Zero ohm standard resistor) |
※ We will change the specifications of the catalogue without notice by improvement.