AE-1652E Wide range, High speed and High accuracy Digital resistance checker 

A wide range from low resistor to high resistor measurement is achieved by the highly precise and wonderful speed
Optimum for the on a sorting machine,taping machine and painting conveyor
For chip MELF and lead typs resistor of B,C,D,F,G,K and M class

  • Ultra high accuracy and high stability by the measuring method rejected thermoelectromotive force.
  • High stability by the improvement of noise immunity for the isolated circuit
    between analog and digital part.
  • Available to make the high speed and ultra-high stable measurement by the setting
    function of average time on measureing value for each range.
  • Range of measurement for absolute value : 0.000mΩ ~ 125.00MΩ
    for % : -9.999% ~ +2.5000% / -99.99% ~ +25.00%
  • Available to select the function for contact check “before measurement”,
    “after measurement” or “non-contact check”
  • 【New functions】Contact check function while measurements. (Always)
       ※Available to improve reliable measurement.
  • RS-232C and Centronics interface are built-in as standard equipment.【GP-IB is option】
  • Transfer function of setting date is built-in as standard equipment.
    (Available to transfer the same setting data to another set of AE-1652E)
  • The checking circuit of the abnormal measuring current and voltage is built-in.
Measurning range and Accuracy(at 23℃±5℃)、90days after colibration 【1year after calibration:2times】

   

Range Standerd setting range Test Current Accuracy
SLOW[Hi-Reso.] FAST[Hi-Reso.] Additional error1 Additional error2
10mΩ 0.500mΩ~10.900mΩ 1A Whith in±0.010%±4×β±10d ±0.010%±4×β±15d±[4/(1+n)]d ±(40×γ) -
100mΩ 5.00mΩ~109.00mΩ 1A
*100mA
Whith in±0.010%±β±5d Whith in±0.010%±2×β±
10d±[4/(1+n)]d
±(4×γ) ±15×(β+2×γ)
0.0500Ω~1.0900Ω 100mA
*10mA
Whith in±0.010%±β±5d Whith in±0.010%±β±10d±
[2/(1+n)]d
±6×(β+3×γ)
10Ω 0.500Ω~10.900Ω 50mA
*5mA
- ±5×(β+γ)
100Ω 5.00Ω~109.00Ω 10mA
*5mA
Whith in±0.010%± β2d Whith in±0.010%±β±10±[1/(1+n)]d -
1kΩ 0.0500kΩ~1.090kΩ 5mA - -
10kΩ 0.500kΩ~10.900kΩ 0.5mA - -
100kΩ 5.00kΩ~109.00kΩ 50μA - -
1MΩ 0.0500MΩ~1.0900MΩ 5μA Whith in±0.025%±β10d±[1/(1+n)]d - -
10MΩ Whith in±0.02%±β±10d 0.5μA ±0.02%±β±10d以内 ±0.1%±β±20d±[1/(1+n)]d以内 - -
100MΩ 5.00MΩ~109.00MΩ 0.05μA Whith in±0.1%±β±20d - - -

d:digits n:program[12]measurement time(msec) in case of absolute varue: apply to LO-Reso(add to ±1d) *Program[18]In case of Lo current mode is ON
in case of %value:α=(100/standerd setting value 10mΩ)×0.01%, α=0
β=(range//standerd setting valu)×0.001% γ=(β/average setting value)program [17]average count

Measurement time external start Free running
SLOW FAST SLOW FAST
about 17sec.~36msec about 0.7msec.~23msec. about30count/sec ~ about 5count/sec About 60 count/sec~about 10 count/sec
Test terminalopen voltage below 15V
EOC[Endofcomparison]pulsewidth 1~250msec. or continuative
Measuring method 2 or 4 terminal measurement
Comparator set range % range:±9.999% / -99.99%~+25.00%
Absolute value:00000~12500
Operation condition temperature:0℃~+50℃、 humidity:80%below
Power supply AC85V~265V、50~60Hz、about 50VA
Outer dimension 333(W)×99(H)×300(D)mm(excluding protruding parts such as rubber legs, etc)
Weight About 3.6kg
OPTION ・GP-IB
・Data cable
・Short termination(Zero ohm standard resistor)

※ We will change the specifications of the catalogue without notice by improvement.

Copyright(c) 2014 AEMIC CORPORATION All Rights Reserved.