“NEW!”AE-182A Ultra-high Speed and High Accuracy DC resistance checker

Optimal for sorting machine and taping of D,F,G,J,K class resistance,inductor and RDC measurement of thermistor

New function

  • New reliability improving method: contact-check function during measurement
  • Help to improve yield by the retry function during measurement
  • To achieve a about 2~2.7 times speedup while keeping the accuracy and substantially accelerating measurement speed for both low and high resistance area
  • Possible to manage measurement data by attached application(for Windows)
    (Data can be saved in csv)
  • LAN interface as option
  • Bar code(QR code)interface as option

Characteristic

  • Corresponding to very small chip resistor(low power measurement)
  • Super high speed(As representative value ,from START to EOC is only 0.9msec according
    to actual measurement)
  • Realize high-accuracy and stable measurement by cancelling thermo electromotive force
  • Realize ultra-high speed・high-stable measurement by the function of setting integral time
    for each range
  • %measurement:-99.99%〜+25.00%【STANDARD setting range:5mΩ~109MΩ】
  • Absolute value setting:0.00mΩ~125.00MΩ
  • Contact check:available to select before measurement/after measurement/OFF(Always
    monitor the contact statement during measurement in each statement)
  • Equipped with contact-cleaning as standard function
  • Equipped with RS-232C interface as standard function (GP-IB/LAN/bar code is option)
  • Equipped with printout as standard function(Centronics conformity)
  • Transfer function for setting data is equipped as standard (Available to automatically set
    and transfer the same setting data to another AE-182A,AE-162E)
  • Equipped with check abnormalities function for measurement current/measurement voltage
    /ADC data as standard
Measurning range and Accuracy(at 23℃±5℃)、180days after calibration【1year after calibration : 1.5times】
Measuring Range Standerd Value setting range Test Current Accuracy
SLOW FAST
100mΩ 5.00mΩ~109.00mΩ 100mA within±0.02%±2α±2d within±0.03%±3α±2d±[2/(1+n)]d
0.050Ω~1.090Ω 100mA within±0.02%±α±1d within±0.02%±α±2d±[2/(1+n)]d
10Ω 0.500Ω~10.900Ω 50mA
100Ω 5.00Ω~109.00Ω 10mA within±0.02%±1d within±0.02%±2d±[1/(1+n)]d
1kΩ 0.050kΩ~1.0900kΩ 5mA
10kΩ 0.500kΩ~10.900kΩ 0.5mA
100kΩ 5.000kΩ~109.00kΩ 50μA
1MΩ 0.050kΩ~1.0900MΩ 5μA within±0.05%±2d±[1/(1+n)]d
10MΩ 0.500MΩ~10.900MΩ 0.5μA within±0.03%±1d within±0.2%±4d±[1/(1+n)]d
100MΩ 5.000MΩ~109.00MΩ 0.05μA within±0.1%±2d
※ d : digit n : Averaging time(msec.), In percent measurement:α=(100/ Standard Setting Value mΩ)×0.01%、In measurement of absolute value:
α=0(add±1d) ※The above accuracy is under the condition shielded completely.
Measurement time External Start Free Running
SLOW FAST SLOW FAST
About 17.4msec.~400msec. About 0.9msec.~400msec. About30~2times/sec. About60~2times/sec.

※ Measurement speed is within accuracy guarantee range. In actual working station, to shorten measurement compared with AE-162E for all ranges

Test terminal open voltage Below 14V
Measurement completion signal(EOC) pulse width 1~250msec. or continuative
Measuring method 2 or 4 terminal measurement capable of switching
Determination value setting range %measurement:-99.99〜+25.00%, absolute value measurement:00000~12500
Operation condition Temp.:0℃~+50℃、 Humidity:below 80%
Power supply AC85V~265V、50~60Hz、about50VA
Outer dimension 250(W)×92(H)×300(D)mm(excluding protruding parts such as rubber legs)
Weight About 2.7kg
AE-182A ・External control/O、RS-232C、Centronics standard
Option ・GP-IB
・LAN
・Bar code unit(reader sold separately)
・Data transfer cable
・Short termination( Zero ohm standard resistor )

※ As to other general-purpose option (such as tweezers probe, measurement terminal convert
connector), please refer to accessory catalogue.