AE-323A High Speed Type・1kHz Digital Capacitor Checker
Optimum for the use on a taping machine for Film Capacitor and Ceramic Capacitor.
- Measuring range:10pF~9.99μF/±99.9%
- 5-Terminal measurement.
- High speed integration method measurement.
- Measuring value is displayed in percentage against the reference capacitance value set
Digitally and comparison result by built-in comparator is displayed as –NG,GO,+NG by LED and buzzer. - BCD parallel out available.
- External DC bias possible(max.DC25V)
Range No. | Measuring range | Test voltage | Accuracy |
---|---|---|---|
0 | 10pF~99.9pF | 1V(r.m.s)approx. | within±0.25%±0.25p±1digit |
1 | 100pF~999pF | ||
2 | 1nF~9.99nF | ||
3 | 10nF~99.9nF | ||
4 | 100nF~999nF | ||
5 | 1μF~9.99μF |
Measuring method: | 5-terminal measuring method、parallel equivalent circuit. |
---|---|
Measuring frequency: | 1kHz±0.1%, sine wave form. |
Stray capacitance cancel range: | 10pF approx. |
Measuring time: | 【in case of remote start mode】about 3.2msec. |
【in case of free running mode】about 68time/sec. | |
Comparator setting range: | Upper limit:0.0%~+99.9%, Lower limit:0.0%~-99.9% |
Display of comparison result by comparator: | LED display of -NG/GO/+NG, and buzzer. |
Comparator setting range: | Upper limit : 0.0% ~ +99.9%, Lower limit : 0.0% ~ -99.9% |
Display of comparison result by comparator: | LED display of -NG/GO/+NG,and buzzer. |
Input and output signal: | Remote start input:0~+12V or open collector. |
Remote start input:0~+12V or open collector. | |
End of measurement output[EOC]:open collector [max.80V/100mA] | |
BCD parallel output of measuring value:TTL fan out 2 | |
Operating condition: | Temp.:0℃~+50℃、 Humidity:below 85% |
Power supply: | AC85V~265V、50~60Hz、about 15VA |
Outer dimension: | 250(W)×89(H)×250(D)mm(excluding protruding parts such as rubber legs, etc.) |
Weight: | 2kg approx. |
※ We will change the specifications of the catalogue without notice by improvement.