“NEW”AE-1152F 5A Ultra Low Ohm METER
1mΩrange measurement current5A
Optimal for measurement of ultra low resistant such as shunt resistor
- 1mΩrange measurement current 5A、possible to measure 1mΩ resistor or less with higher precision and improve the accuracy by 0.30% compared with AE-1152D.
- %measurement:STANDARD setting range 0.01mΩ~1kΩ
:measurement range ±50.00%【minimum resolution 1nΩ】 - Absolute value measurement:0.0000mΩ~1.5000kΩ
- Possible to cancel the influence of thermal electromotive force.
- Measurement current is applied in pulse to help reduce heating of a target object and wear of measurement terminal .
- Equipped with contact check function as standard.
- Possible to add various option(only one of those can be chosen).
GP-IB / RS-232C / LAN / Centronics interface /
Bar code/ two-dimensional code reader +RS—232C interface
And data collection application for RS-232C is also attached. - Result of comparator can be outputted in open collector output .LED display and buzzer sound are also possible.
- Measurement current / measurement voltage check circuit is equipped as standard.
- Available for shift output.
Measuring range and Accuracy (at 23℃±5℃) :
| Range | Measuring range | Resolution | Current | Accuracy | |
|---|---|---|---|---|---|
| Absolute value measurement | %measurement | ||||
| 1mΩ | 0.0000mΩ~1.5000mΩ | 0.1μΩ | 5A | Within±(0.01%rdg+0.3μΩ) ±3 digit [Average]/ ±4 digit [Slow]/ ±5 digit [Fast] |
|
| 10mΩ | 0.000mΩ~15.000mΩ | 1μΩ | 1A | ||
| 100mΩ | 0.00mΩ~150.00mΩ | 10μΩ | |||
| 1Ω | 0.0000Ω~1.5000Ω | 100μΩ | 100mA | ||
| 10Ω | 0.000Ω~15.000Ω | 1mΩ | |||
| 100Ω | 0.00Ω~150.00Ω | 10mΩ | 10mA | ||
| 1KΩ | 0.0000kΩ~1.5000kΩ | 100mΩ | 1mA | ||
| % | 0.01mΩ~1kΩ/±50.00% | 0.01%【1nΩ】 | Refer to above | ||
※In% measurement, accuracy of 10mΩ range or less applies in double ON and should be the
reference value in double OFF.
※α:SLOW・AVERAGE=3/SLOW=4/FAST=5
β:1mΩrange=0.2/other range=1
| Voltage at terminals open : | below 4V |
|---|---|
| Measuring method : | 4-terminal measurement [contact check is available] |
| Sampling time : | 【Free running mode】2~10times/sec. |
| 【External start】about 9msec~400msec. | |
| Comparator setting range : | Absolute value measurement :0~15000[both for low and high limit],%measurement: low limit -50〜0%,high limit :0~+50% |
| Indication of comparator determination result : | LO/GO/HI, LED display and buzzer. |
| Control signal : | Start signal:start by ”L”[0V]→”H”[DC12V] |
| Hold signal:Open and”H”[DC12V]:Free run,”L”[0V]:Hold | |
| Determination result signal[LO/GO/HI]: open collector output max40V、100mA | |
| Contact check error [CONT-E]: open collector output max40V、100mA | |
| Termination signal[EOC]: open collector output max40V、100mA | |
| Operation condition : | 【Temp.】5℃~+40℃、【Humidity】below 85% |
| Power supply : | AC85V~265V、50~60Hz、about 60VA |
| Outer dimension : | 333(W)×99(H)×300(D)mm【excluding protruding parts such as rubber legs, etc.】 |
| Weight : | about 3kg |
| Option | ・RS-232C I/F | Only one kind of interface can be provided with the checker. |
|---|---|---|
| ・GP-IB I/F | ||
| ・LAN I/F | ||
| ・Bar code/ two-dimensional code reader +RS—232C I/F | ||
| ・Printer output (Centronics)I/F | ||
| ・Each cable(RS-232C,GP—IB,printer) | ||
| ・Short termination(Zero ohm standard resistor) | ||