AE-3122A High Speed Type・120Hz Digital Capacitor Checker

Optimum for the use on a taping machine for Aluminum Electrolytic Capacitor and Tantalum Capacitor.

  • Measuring range:100nF~9.99mF/-79.9%~+99.9%
  • 4-Terminal measurement.
  • High speed integration method measurement.
  • Measuring value is displayed in percentage against the reference capacitance value set
    Digitally and comparison result by built-in comparator is displayed as –NG,GO,+NG by LED and buzzer.
  • BCD parallel out available.
  • External DC bias possible(max.DC25V)
Range No. Measuring range Test voltage Accuracy
1 100nF~999nF within 100mV(r.m.s) within±0.25%±1digit
2 1μF~9.99μF
3 10μF~99.9μF
4 100μF~999μF
5 1mF~9.99mF within±0.5%±3digit
Measuring method: 4-terminal measuring method、serial equivalent circuit.
Measuring frequency: 120Hz±0.1%, sine wave form.
Measuring time: 【in case of remote start mode】about 25msec.
【in case of free running mode】about 37.5time/sec.
Display of measured value: -79.9%~+99.9%
Comparator setting range: Upper limit:0.0%~+99.9%, Lower limit:0.0%~-79.9%
Input and output signal: Remote start input:0~+12V or open collector.
Comparison result output[LO/GO/HI]:open collector [max.80V/100mA]
End of measurement output[EOC]:open collector [max.80V/100mA]
BCD parallel output of measuring value:TTL fan out 2
Operating condition: Temp.:0℃~+50℃、 Humidity:below 85%
Power supply: AC85V~265V、50~60Hz、about 20VA
Outer dimension: 250(W)×89(H)×250(D)mm(excluding protruding parts such as rubber legs, etc.)
Weight: 2kg approx.

※ We will change the specifications of the catalogue without notice by improvement.